Weidmuller PT 100/4 signal conditioners for voltage output Manual do Utilizador Página 5

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 14
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 4
19
4.2 Basic calibration with a signal output from 0 ... 10 V
1. Select the temperature range on the printed circuit board
using the DIP switch, see table on the module or pages 23
and 24.
The DIP switches 1, 2 and 3 set the minimum input tempera-
ture ϑ
min
.
The span (difference between minimum and maximum input
temperature) is set using the DIP switches 4, 5 and 6.
2. Professionally install module.
3. Add 1 % of the temperature span to the selected “minimum
temperature and set this value on a PT 100 simulator (when
using a precision resistance decade, where necessary ob-
serve DIN IEC 751 conversion table from °C to !) and cali-
brate the signal conditioner output signal to 0.100 V using the
null potentiometer. (The potentiometers are located behind
the hinged cover).
4. Set the “maximum” selected temperature on a PT 100 simu-
lator (or precision resistance decade) and calibrate the signal
output to 10.000 V using the Span potentiometer.
5. Repeat steps 3 and 4 (approx. 2-3 times), until the required
accuracy is achieved.
Vista de página 4
1 2 3 4 5 6 7 8 9 10 ... 13 14

Comentários a estes Manuais

Sem comentários