20
• Calibration
The PMX400/TMP is factory calibrated. Once calibrated the PMX400/
TMP will operate with any sensor type or range, and any output type
or range, without recalibration. You should not have to recalibrate the
instrument before installation.
Equipment requirements
• A suitable millivolt source (accurate to 5 μV or better)
• An accurate digital multimeter (accurate to 0.05 mV and ±0.1 μA)
• A Pt100 Rtd simulator (accurate to 0.05 ˚C)
Terminal Connections for calibration
Calibration Stage Signal type Terminal
milliVolt inputs
Millivolt source (+) 11
Millivolt source (–) 10
CJC Circuit
Millivolt source (+) 9
Millivolt source (–) 10
RTD Inputs
Rtd simulator B 10
Rtd Simulator B
sense
11
Rtd Simulator A 12
Analogue Current Output
mA output + 3
mA output – 4
Analogue Voltage Output
V output + 3
V output – 5
Summary
There are four stages to calibrating the PMX400/TMP
• Calibrate the millivolt inputs
• Calibrate the CJC input
• Calibrate the RTD input
• Calibrate the analogue outputs